The Xenogen IVIS-200 System is a high-sensitivity, low noise, in vivo imaging technology platform that allows non-invasive visualization and tracking of cellular and genetic activity within a living organism, in real time. It’s capable of imaging bioluminescence and fluorescence. IVIS-200 provides Living Image® software for image display and analysis.
Instrument Specifications
- CCD Cameras
- Sensor
- Back Illuminated
- CCD Format
- 2048 x 2048 pixels
- CCD Size
- 26 x 26 mm
- Effective CCD Format
- 1920 x 1920 pixels
- Pixel Size
- 13.5 x 13.5 um
- Quantum Efficiency
- ~85% 400-700 nm
- >50% 350-900 nm
- Readout Noise
- <5 electrons RMS
- Dark Current
- <100 electrons/s/cm2 (-105°c)
- Minimum Detectable Luminance
- <70 photons/s/sr/cm2
- CCD Temperature
- Nominal ~105°C
- ~85% 400-700 nm
- Optics
- Lens f/stop
- f/1~f/8
- Field of views (FOV)
- 3.9, 6.5, 13, 19.5 and 26 cm
- Resolution
- >60 um (FOV=3.9, f/1)
- Scanning Laser
- <1 mW at 532 nm
- Lens f/stop
- Fluorescence Imaging Components
- Excitation Filters
- 6 positions, 25 mm diameter
- Emission Filters
- 22 positions, 60 mm diameter
- Background (Autofluorescence and Leakage)
- 1 part in 107 typical
- Lamp
- 150 W Quartz Halogen 21 V 3250° Kelvin
- Excitation Filters
There are fluorescence filter sets available on the system are as follows:
Filter Set
Label
Excitation
Emission Passband
Dyes & Passband
1 GFP 445-490 515-575 GFP, EGFP, FITC 2 DsRed 500-550 575-650 DsRed2-1, PKH26, CellTrackerOrange 3 Cy5.5 615-665 695-770 Cy5.5, Alexa Fluor® 660, Alexa Fluor® 680 4 ICG 710-760 810-875 Indocyanine green (ICG) - Sensor